Radiation Physics Equations
Diagnostic Radiology
Film
, where OD is optical density, is amount of incident light, and is amount of transmitted (measured) light
OD values are additive
H and D curve (Hurter-Driffield) gives relationship between OD and absorbed dose. Sigmoid shape
Flat region: OD independent of dose
Toe region: OD increases rapidly
Linear region: OD increases linearly with dose
Saturation region: OD doesn't increase as function of dose
Photon Dosimetry
Atomic coefficient dependence
Note: Probability of interation is not the same as mass attenuation coefficient
Consult Page 36-39 of IAEA text (radiation oncology physics)
Below are the Mass attenuation coefficient dependencies
Coherent scattering ≈ Z
Photoelectric absorption ≈ Z3/E3
Compton scattering ≈ independent of Z, ≈ 1/E, ≈ electrons/gram
Pair production ≈ Z
Triple production ≈ Z2
Hounsfield units
HU = 1000* (μtissue - μwater) / μwater
Heterogeneity corrections
Lung: 10 cm of lung ≈ 3 cm of tissue = 3.3x
Bone: 10 cm of bone ≈ 16 cm of tissue = 0.6x
With higher energy, less correction necessary (since Compton effect is 1/E)
With higher energy, slower build-up at lung/tumor interface, and thus possibly underdosing
If no correction, higher dose at prescription point due to lower attenuation in lung
LET
Specific ionization: number of ion pairs formed per unit path length; depends on velocity and particle charge
Energy transferred to medium per unit path length (energy gain)
LET is proportionate to (Q2 * ρ) / (v2 * Z)
LET = Specific ionization * W
Stopping power
Energy deposited by particle; depends on charge and density of medium
Colisional: lost due to collisional processes (secondary electrons); predominates, especially at lower energies
Radiative: lost due to radiative processes (photons, high energy secondary electrons)
Restricted stopping power: energy lost by particle per unit length, locally absorbed
Inverse square law: I2/I1 = (r1/r2)2
Back scatter factor (SSD setup): BSF = Exposure at surface / Exposure in air
Dose = Exposure (X) * f * BSF
Only applies at low energies, dmax at surface
Peak scatter factor (SSD setup): PSF = Dose at dmax / Dose in air
d_max
Photon d_max (cm)
Co-60 0.5
4MV 1.0
6MV 1.5
10MV 2.5
15MV 3.0
18MV 3.2
20MV 3.5
25MV 4.0
In most centers, we have 6MV, 10MV and 18MV so
6MV : 1.5cm
10MV : 2.5cm
18MV : 3.2cm
Photon attenuation
Co-60 ~4.0% per 1 cm depth
6MV ~3.5% per 1 cm depth
20MV ~2.0% per 1 cm depth
PDD
Percent depth dose (SSD setup): PDD = Dose at depth / Dose at dmax
Two components: patient attenuation and inverse square dose fall-off
Factors that affect PDD:
Energy ==> Increases
Field size ==> Increases
SSD ==> Increases
Depth ==>Decreases
D2 = D1 * (PDD2 / PDD1)
By energy at 100 cm SSD, 10x10 field, and depth of 10cm
Co-60 56%
4MV 61%
6MV 67%
10MV 73%
20MV 80%
25MV 83%
Equivalent squares
Square area that has the same PDD as the rectangular field
--- This is only true for W = L since
Otherwise:
.
See, The Physics of Radiation Therapy by Khan, Chapter 9, p. 185.
Equivalent Square for circular field (D=diameter)
See reference .
A square with side a will be equivalent to a circle with radius r when they have the same area, , so , or
Elliptical fields:
Equivalent diameter of elliptical fields:
-- see 15507419
Skin dose
Factors that affect Skin dose:
Energy ==> Decreases
SSD ==> Decreases
Field size ==> Increases
Bolus ==> Increases
Oblique incidence ==>Increases
Dose Ratios
Mayneord F-factor:
Tissue air ratio (SAD setup): TAR = Dose at depth / Dose in air
Tissue phantom ratio (SAD setup): TPR = Dose at depth / Dose at reference depth
Tissue maximum ratio (SAD setup): TMR = Dose at depth / Dose at dmax
via inverse square correction
MU Calculation
Treatment time or monitor units:
where OF is the output factor, WF is the wedge factor, TF is the tray factor, and ISF is the inverse square factor.
Wedges
Wedge angle: angle by which the isodose curve is turned by the wedge, typically at 10 cm
Hinge angle: angle between the central axes of two incident beams
Dose for arbitrary wedge field θ using flying wedge or dynamic wedge = W0*dose0 + W60*dose60, where W0 = 1-W60, and W60 = tan θ/tan 60
Penumbra
P = s * (SSD + d - SDD) / SDD, where s is source width and SDD is source-diaphragm/collimator distance
Superficial energies
HVL (in Al or Cu) specifies penetrability of low-energy photon beam. HVL is determined by the combination of kVp and filtration (different combinations can give same HVL)
Typically short SSD is used
Compared with electrons, superficial photons have sharper penumbra, deliver higher skin dose, but also higher dose to underlying tissues
Blocks
Dose under 1.5 cm width block (5 HVL), in 15 x 15 cm field, 6 MV, 5 cm depth is ~15% of open field dose. Transmitted dose is ~3% (shielded by 5 HVL), scattered dose from open field contributes the rest
Scattered dose
Patient with pacemaker, if dose to pacemaker to be <5%, need to be at least 2cm from 6 MV beam edge
Patient with breast tangents, ovaries 20 cm from field: dose to ovaries ~0.5%
Dose at 1 m laterally from treatment beam: ~0.1%
Treatment margins
PTV margin
PTV margin = 2.5 (quadratic sum of standard deviation of all preparation (systematic) errors) + 0.7 * (quadratic sum of standard deviation of all execution (random) errors) 10863086 (2000: van Herk M, Int J Radiat Oncol Biol Phys. 2000 Jul 1;47(4):1121-35.)
PTV margin = 2.5 sigma + 0.7 delta (cover CTV for 90% of patients with 95% isodose)
Electron Dosimetry
Probability of bremsstrahlung interaction: Z2
X-ray emission spectrum proportionate to kVp2 * mAs / d2, also depends on amount of filtration
Lead block thickness to attenuate 95%: tPb (mm) = Electron energy / 2
Cerrobend block thickness tCerr = 1.2 * tPb
Range
Practical range in water: Rp (cm) = Electron energy / 2
R50: depth at which dose is 50% of maximum
Depth of calibration
I50: Find depth of 50% ionization in water
R50: Calculate R50 = 1.029 * I50 - 0.06 if <10 cm depth, R50=1.059 * I50 - 0.37 if >10 cm depth
dref = 0.6 * R50 - 0.1
Energy is specified by the R50 parameter
Typically treated as SSD setup
No physical source in accelerator head; clinical beams appears to emerge from a "virtual source". Can be found by backprojecting beam profiles at different depths
Virtual SSD shorter than actual (photon) SSD
Inverse square corrections can be done on virtual SSD for large fields; for small fields effective SSD should be determined
Output Dose rate = Applicator Dose rate * Back scatter factor(cutout)/Back scatter factor(Applicator)/ (SSD/SSD+SO)^2 (SSD= Source to surface distance & SO= Stand Off)
Radiation Quality
Half Value Layer: HVL = ln 2 / μ
Tenth Value Layer: 1 TVL = 3.32 HVL
Attenuation: N = N0 * e-μx, where N is number of photons remaining, μ is linear attenuation coefficient, x is thickness of block
Attenuation: N = N0 * (1/2)n, where n is number of HVLs
Brachytherapy
1 Ci = 37 x 109 Bq
Activity: A = A0 * e-λt
Activity: A = A0 * (1/2)n, where n is number of half-lives elapsed
Specific activity: SA = A / m = λ * (Na / AW)
Half-life: t1/2 = ln 2 / λ
Mean (average) life: tavg = 1 / λ = 1.44 * t1/2
Permanent implant: Dosetotal = Dose rate0 * tavg
Temporary implant: Dosetotal = Dose rate0 * tavg * (1 - exp(-t/tavg) = Dose rate0 * tavg * (1 - exp(-λt))
Exposure rate: X = Γ * Α / d2
Where Γ is gamma constant, A is activity, and d is distance from source
Dose rate: D = Sk * Λ * G * F * g
Where Sk is air-kerma strength, Λ is dose-rate constant, G is geometry factor (see below), F is anisotropy factor, and g is radial dose function
Geometry factor G(r,θ)
Point source: 1/r2
Line source: (θ2 - θ1)/Ly, where L is length of line, y is distance
ICRU dose rate:
Low 0.4 - 2.0 Gy/h
Medium 2.0 - 12.0 Gy/h
High >12.0 Gy/h
Brachytherapy systems
Paterson-Parker (Manchester): non-uniform needles (1/3, 1/2, 2/3 center vs periphery depending on plane size), uniform dose
Quimby: uniform needles, non-uniform dose (higher in center)
Shielding
Workload (W): Beam-on time (in Gy at 1 m from source)
Use factor (U): Fraction of time beam aimed at particular target (dimensionless)
Occupancy factor (T): Fraction of time area is occupied by an individual (dimensionless)
Distance (d): from isocenter to area of interest (m)
Barrier transmission factor (B): amount of radiation passing through barrier
Permissible dose (P): maximum dose for an area of interest (Gy)
Shielding equations
Primary barrier dose equation:
Primary barrier shielding equation:
Secondary barrier scattering equation:
where α is the scattered fraction, diso is the distance from the source to the isocenter, dwall is the distance from the isocenter to the wall, and F is the maximum field area in cm2.
Secondary barrier leakage equation:
where dhead is the minimum distance from the linac head to the wall.
Internal Sources
Effective half-life: Accounts for physical half-life and for biologic half-life, always less than either
teff,uptake = (tbiol, uptake * tphys) / (tbiol, uptake + tphys)
teff,elim = (tbiol, elim * tphys) / (tbiol, elim + tphys)
Radiation Protection
Dose equivalent (H): Absorbed dose (D) * WR * N
WR, previously known as Q, is the quality factor
N is geometry factor
Unit in Sievert (Sv)
Effective dose equivalent (HT): Sum of H for a given tissue across different radiation types (e.g. for nuclear explosion)
Formerly known as "equivalent" dose
Effective dose (E): Sum of HT for whole body across different tissues
Gonads have WT = 0.12 (lower than lung/breasts/stomach/bone marrow/colon)